Products

Carbolite PF Fan Convection Oven

Laboratory ovens of the Peak series 300°C are available as PF fan convection oven(s) and PN natural convection oven(s).

  • Fan oven: Forced convection
  • Max temp: 250 – 300 °C
  • Min temp: Ambient +30°C
  • Volume: 28 to 910 litres

Laboratory ovens of the Peak series 300°C are available as PF fan convection oven and PN natural convection oven models.

FP fan convection oven provides greater temperature uniformity and faster recovery rates than natural convection.

Standard Features

  • 300°C (PF 30 to PF 200) or 250°C (PF 400, PF 800) maximum operating temperatures
  • 28 to 910-litre chamber volumes
  • Equipped with the R38 digital PID temperature controller (PF 30 to PF 200)
  • Carbolite Gero 301 controller, with single ramp to set-point & process timer (PF 400 & PF 800)
  • Fan convection for rapid heating & recovery & excellent uniformity
  • Chemically resistant stainless steel liner
  • Two nickel-chrome plated wire shelves (The PF 400 is supplied with 3 wire shelves, the PF 800 with 3 perforated stainless steel shelves)
  • Lever latch door & airtight silicone seal
  • Compliant with safety standard BS EN 61010-2-010:2003
Options (SPECIFY THESE AT THE TIME OF ORDER)
  • A range of sophisticated digital controllers, multisegment programmers and data loggers with digital communication options is available – more information about controllers
  • Over-temperature protection (recommended to protect valuable contents & for unattended operation)
  • Access port for independent thermocouple
  • Accessory shelves & runners
  • Heavy-duty reinforced base and shelves
  • Cable entry ports
  • Variable speed fan control
  • Stoving & curing options for the extraction of small volumes of volatile solvents (not compatible with the viewing window option)
  • Viewing window (not compatible with stoving & curing option)
  • Interior light (not compatible with stoving & curing option)
  • Air exhaust fan
  • Moisture extraction option (comprising sealed seams and air exhaust fan)
  • Stacking frame to enable units to be stacked one upon another
  • Lockable door
  • Door switch to isolate elements and fan
  • Fixed or castor-mounted floor stands
  • Low voltage options for use below 220 V
  • Routine spares kit

Technical Details (Models)

PF30

Max temp (°C) 300
Min temp (°C) Ambient +30
Volume (litres) 28
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 300°C
Heat-up time (mins) 40
Recovery time (mins) 4
Dimensions:
Internal H x W x D (mm)
300 x 290 x 320
Dimensions:
External H x W x D (mm)
470 x 665 x 470
Shelves fitted/accepted 2 / 3
Shelf loading each / total (kg) 10 / 20
Optional heavy-duty reinforced base max. weight (kg) 40
Optional heavy-duty shelves max. quantity 2
Optional heavy-duty shelves max. weight/shelf (kg) 20
Configuration Bench-top
Air changes/hr 50 / 360*
Max power (W) 1000
Holding power (W) 560
Weight (kg) 37

PF60

Max temp (°C) 300
Min temp (°C) Ambient +30
Volume (litres) 66
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 300°C
Heat-up time (mins) 36
Recovery time (mins) 4
Dimensions:
Internal H x W x D (mm)
400 x 390 x 420
Dimensions:
External H x W x D (mm)
570 x 765 x 570
Shelves fitted/accepted 2 / 5
Shelf loading each / total (kg) 10 / 30
Optional heavy-duty reinforced base max. weight (kg) 60
Optional heavy-duty shelves max. quantity 3
Optional heavy-duty shelves max. weight/shelf (kg) 20
Configuration Bench-top
Air changes/hr 21 / 153*
Max power (W) 1500
Holding power (W) 775
Weight (kg) 55

PF120

Max temp (°C) 300
Min temp (°C) Ambient +30
Volume (litres) 127
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 300°C
Heat-up time (mins) 35
Recovery time (mins) 4
Dimensions:
Internal H x W x D (mm)
500 x 490 x 520
Dimensions:
External H x W x D (mm)
670 x 865 x 670
Shelves fitted/accepted 2 / 9
Shelf loading each / total (kg) 10 / 40
Optional heavy-duty reinforced base max. weight (kg) 80
Optional heavy-duty shelves max. quantity 3
Optional heavy-duty shelves max. weight/shelf (kg) 25
Configuration Bench-top
Air changes/hr 11 / 79*
Max power (W) 2000
Holding power (W) 900
Weight (kg) 74

PF200

Max temp (°C) 300
Min temp (°C) Ambient +30
Volume (litres) 230
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 300°C
Heat-up time (mins) 42
Recovery time (mins) 5
Dimensions:
Internal H x W x D (mm)
750 x 590 x 520
Dimensions:
External H x W x D (mm)
920 x 965 x 670
Shelves fitted/accepted 2 / 15
Shelf loading each / total (kg) 10 / 50
Optional heavy-duty reinforced base max. weight (kg) 100
Optional heavy-duty shelves max. quantity 4
Optional heavy-duty shelves max. weight/shelf (kg) 25
Configuration Bench-top
Air changes/hr 6 / 44*
Max power (W) 2700
Holding power (W) 1180
Weight (kg) 96

PF400

Max temp (°C) 250
Min temp (°C) Ambient +30
Volume (litres) 460
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 250°C
Heat-up time (mins) 85
Recovery time (mins) 25
Dimensions:
Internal H x W x D (mm)
1500 x 605 x 510
Dimensions:
External H x W x D (mm)
1835 x 1025 x 1100
Shelves fitted/accepted 3 / 14
Shelf loading each / total (kg) 10 / 75
Optional heavy-duty reinforced base max. weight (kg) 150
Optional heavy-duty shelves max. quantity 5
Optional heavy-duty shelves max. weight/shelf (kg) 25
Configuration Floor-standing
Air changes/hr 30
Max power (W) 6000
Holding power (W) 2200
Weight (kg) 200

PF800

Max temp (°C) 250
Min temp (°C) Ambient +30
Volume (litres) 910
Temp stability (°C) ±0.2
Temp uniformity (°C) ±5.0 @ 250°C
Heat-up time (mins) 100
Recovery time (mins) 30
Dimensions:
Internal H x W x D (mm)
1500 x 1200 x 510
Dimensions:
External H x W x D (mm)
1835 x 1615 x 1100
Shelves fitted/accepted 3 / 7
Shelf loading each / total (kg) 10 / 100
Optional heavy-duty reinforced base max. weight (kg) 225
Optional heavy-duty shelves max. quantity 5
Optional heavy-duty shelves max. weight/shelf (kg) 75
Configuration Floor-standing
Air changes/hr 15
Max power (W) 9000
Holding power (W) 3500
Weight (kg) 280

Please note
– Uniformity is measured in an empty chamber with vents closed after a stabilisation period
– Shelf loadings are based on evenly distributed weight
– The uniform volume is smaller than the total chamber volume
– Maximum power and heat-up times based on a 240V supply
– Stoving and curing options may require increased maximum power
– * When equipped with an optional exhaust fan

Content may be subject to modifications or corrections

Subscribe

Known for service-excellence, we don’t just supply the tools, we assist you in gaining proficiency with our products through our dedicated training program.

Scroll to Top