Products

Carbolite TLD Rapid Cooling Oven

The TLD rapid cooling ovens are frequently used for annealing thermo-luminescent dosimeters (TLD) that have been used to measure exposure to ionizing radiation.

  • Max temp: 400 °C
  • Volume: 3 litres

The TLD rapid cooling ovens are frequently used for annealing thermo-luminescent dosimeters to measure exposure to ionizing radiation.

The TLD rapid cooling ovens are designed to heat to 400 °C, cooling rapidly to ambient temperature using forced air cooling. This rapid cycling capability is also suitable for other small-scale tempering and annealing applications.

Standard Features

  • 400°C maximum operating temperature
  • CC-T1 programmable controller providing automatic activation of the cooling blower
  • Horizontal forced air circulation from rear rear-mounted fan
  • Excellent performance & reliability
  • Stainless steel liner
  • Stainless steel mesh shelves
Options (SPECIFY THESE AT THE TIME OF ORDER)
  • A range of sophisticated digital controllers, multisegment programmers and data loggers with digital communication options is available – more information about controllers
  • Independent over-temperature protection with digital setpoint & display
  • Digital process timer

Content may be subject to modifications or corrections

Technical Details (Models)

TLD/3                 TLD/28

Max temp (°C) 400 400
Min temp (°C) 50 50
Temp stability (°C) ±1 ±1
Temp uniformity (°C) ±5.0 ±5.0
Heat-up time (mins) 60 60
Heating/cooling rate (°C/mins) 4* 4*
Dimensions:
Internal H x W x D (mm)
150 x 150 x 100 305 x 305 x 305
Dimensions:
External H x W x D (mm)
530 x 370 x 500 880 x 675 x 865
Shelves fitted/accepted 2 / 2 2 / 2
Shelf loading each / total (kg) 1 / 2 10 / 20
Volume (litres) 3 28
Weight (kg) 26 95
Max power (W) 1000 2250

Please note
– Uniformity is measured in an empty chamber with vents closed after a stabilization period
– Shelf loadings are based on evenly distributed weight
– *Based upon cooling an empty chamber
– The uniform volume is smaller than the total chamber volume
– Minimum operating temperature is ambient +50°C

Content may be subject to modifications or corrections

Subscribe

Known for service-excellence, we don’t just supply the tools, we assist you in gaining proficiency with our products through our dedicated training program.

Scroll to Top